qPCR sealing film is not ordinary PCR film. It has to stay optically clear through 40 thermal cycles, add almost no fluorescence of its own, and hold 20 µL of reaction at 95 °C without losing volume. Five checks decide whether a film is fit for quantitative work — and four of them can be done in the lab with water, a printed page and a centrifuge.
1. Visual screening
- Surface. A good film is smooth and evenly transparent, with no white spots, crystalline patches, pinholes or scratches. Dull, cloudy film with visible particles will scatter light.
- Thickness. Stack several sheets and compare. Uneven thickness means some areas will seal poorly and others will tear during application.
- Odour. Film should smell of nothing much when the pack is opened. A sharp plastic or solvent smell suggests residual volatiles, and those become contamination in a 20 µL reaction.
2. Optical performance — the qPCR differentiator
This is where sealing film for qPCR separates from general-purpose film, because the detector has to read fluorescence through the film.
- Transmittance. Lay the film over printed text of standard size. If the text reads clearly through it, transmittance is very likely 90 % or better. If it does not, the film will attenuate the signal.
- Autofluorescence. Illuminate the film directly. Visible blue or green glow means strong autofluorescence and the film is unsuitable. A more rigorous version: seal a blank plate and run a baseline scan — if the background reads more than 20 % above an accepted control film, reject it.
- Signal stability. The same film should return comparable signal at cycle 10 and cycle 40; a drop across the run points to scattering rather than attenuation.
3. Sealing and temperature resistance
- Evaporation. Add 20 µL of buffer to every well of a 96-well plate, seal, and run 40 standard qPCR cycles. A good film averages 3 % or less evaporation, which is what makes 5–20 µL reaction volumes viable.
- Seal integrity after cycling. After the run, check for blistering, delamination or lifted edges. Invert the plate for 30 seconds with the wells pierced: no leak means the seal held.
- Spin resistance. The film should survive 2,000 ×g for 10 minutes without rupturing.
- Freeze–thaw. Store a sealed plate at -80 °C for 7 days, then thaw. No brittle cracking, no delamination and no condensation inside the wells. Across the two tests the working range is roughly -80 °C to 120 °C.
4. Handling and fit
- Peel strength. 0.5 N/cm or better, holding through high-temperature cycling and peeling off without adhesive residue on the plate rim.
- Puncture. A tip should pass through cleanly without shedding film particles into the well — particles are a contamination route, not just a nuisance.
- Footprint. SBS-standard dimensions, typically 140 × 80 mm for a 96-well plate.
5. Documentation and material
Ask for the test report: DNase, RNase and pyrogen results should be stated, not implied. On material, polyolefin and modified polypropylene give better optical purity than standard PET or PP film, which is the practical difference between a film that reads cleanly and one that adds background.
Pass criteria at a glance
| Check | Pass criterion |
|---|---|
| Light transmittance | 90 % or better (standard printed text legible through the film) |
| Autofluorescence | Baseline no more than 20 % above an accepted control film |
| Evaporation, 40 cycles at 20 µL | 3 % average or less |
| Seal after cycling | No blistering, delamination or lifted edges; inverted 30 s, no leak |
| Spin resistance | 2,000 ×g for 10 min without rupture |
| Freeze resistance | -80 °C for 7 days, no brittleness or delamination |
| Peel strength | 0.5 N/cm or better, no adhesive residue |
| Footprint | SBS standard, 140 × 80 mm |
Frequently asked questions
Can I use ordinary PCR film for qPCR?
Not if you intend to quantify. Standard film is often made for optical clarity at lower demands, and its autofluorescence can swamp a small signal. If the film was not specified for qPCR, assume it will add background.
How do I test a film without running a full plate?
Two quick checks: read printed text through it to judge transmittance, and scan a sealed blank plate on the instrument to measure background against a control film.
Does the film matter at 5–20 µL volumes?
More than at larger volumes. A 3 % evaporation loss is negligible in a 100 µL reaction and significant in a 5 µL one, which is why the evaporation test is run at 20 µL.
Related reading
Note: The information on the scientific research products above is for reference only!
Related BUNSEN products
- qPCR Sealing Films
- PCR Plates & PCR Tubes
- 96-Well Deep Well Plates
- Product Center — all BUNSEN lab consumables & instruments
Browse the full product range, or request a sample or quotation.
Leave a Reply